一、Mobile X-ray machine features:
1.For hospital wards and emergency room photography dedicated
2.Flexible mobile operating performance
3.Wireless remote exposure, greatly reducing the dose of physician radiation
二、The main technical parameters:
1.Power supply conditions:
voltage: communicate with 220V±22V;
Power frequency: 50Hz±0.5Hz;
battery capacity: ≥5kVA;
Power supply allows maximum internal resistance: 1Ω
2、 Photography conditions: Tube voltage: 40-90kV a total of 11 files, each file 5kV;
Tube current: 30,50mA total 2 files;
time : 0.05s-6 s a total of 20 files.
3 、Mechanical behavior:
When the X-ray generator window down, the focus and the distance from the ground range of activities 450mm ~ 1770mm
The X-ray generator has a rotation angle of ± 180º around the horizontal axis
The X-ray generator has a rotation angle of ± 90º around the vertical axis
4 、Overall dimensions: 1290 (length) × 630 (width) × 1935 (high) (mm)